MSK-ESC-R2R Roll to Roll Electrospinning System with Syringe Pump and Heating Vacuum Bed for Uniform Nanofiber Coating
The MSK-ESC-R2R system is engineered for laboratory-scale continuous electrospinning processes, enabling uniform nanofiber deposition on flexible substrates. It integrates an automatic roll to roll substrate feeding and retrieval mechanism, facilitating uninterrupted coating workflows. This setup is ideal for applications requiring precise nanofiber layering with controlled thickness and morphology.
Central to its operation is a digital syringe pump that delivers a consistent flow of polymer solutions, supporting syringe volumes up to 200 mL. The system’s heating vacuum bed, capable of reaching 100°C, aids in solvent evaporation and coating bake-out directly on the substrate. The high voltage power supply offers adjustable output from 0 to 30 kV with a maximum current of 1 mA, ensuring stable electrospinning jets across multiple needles.
Wichtigste Merkmale
- Linear array of 5 syringe needles spaced 25 mm apart for uniform coverage up to 160 mm width.
- Automatic roll to roll system with precise speed control, supporting substrate widths up to 200 mm.
- Heatable vacuum bed with PID temperature control up to 100°C for efficient coating drying.
- Digital syringe pump with multiple operating modes and flow rates from 1.963 μL/min to 294.524 mL/min depending on syringe size.
- Single high voltage supply cable minimizes jet repulsion between needles, enhancing coating uniformity.
- Removable spraying heads accommodate various needle gauges (22Ga to 30Ga) for flexible coating parameters.
- High voltage power supply with digital voltage and current display, constant current and voltage modes, and wide AC input range.
Technische Daten
| Parameter | Spezifikation |
|---|---|
| High Voltage Output | 0 – 30 kV, max 1 mA |
| Voltage/Current Control | Knobs with digital display; constant current and voltage modes |
| Syringe Pump Capacity | Up to 200 mL syringes |
| Pump Flow Rate Range | 1.963 μL/min (10 μL syringe) to 294.524 mL/min (200 mL syringe) |
| Pump Plunger Speed | 0.0001 cm/min to 15 cm/min |
| Needle Configuration | 5 needles per head; needle spacing 25 mm; needle gauges 22Ga, 23Ga, 25Ga, 27Ga, 30Ga |
| Heating Vacuum Bed Size | 600 mm (L) x 300 mm (W) |
| Heating Bed Temperature | Up to 100°C with ±1°C PID control |
| Roll to Roll Width | Up to 200 mm |
| Roll to Roll Speed Range | 2.2 mm/s to 117.2 mm/s (1 to 41 rpm motor speed) |
| Power Requirements | HV device: 24 VDC input; Vacuum bed: 208-240 VAC, 50/60 Hz; Roll to roll: 110 VAC |
Typische Anwendungsbereiche
This system is suited for research and development of nanofiber membranes, air filters, battery separators, and other functional coatings requiring uniform electrospun layers on flexible substrates. It supports polymer solution electrospinning with controlled deposition and in-situ drying.
Hinweise zur Konfiguration und zum Betrieb
The system includes adjustable needle stands and removable spraying heads for flexible coating widths. Operators should use the system within a fume hood to ensure safe handling of volatile solvents. Optional needle gauge sets allow customization of fiber diameter and coating characteristics. The single voltage supply cable design reduces electrostatic interference between needles, improving coating uniformity.
Maße & Versand
| Product dimensions | 300mm (L) x 270mm (W) x 120mm (H) |
|---|