OTF-1700X-UL Split Tube Furnace with 60mm Alumina Tube, Vacuum Flanges, Valves, and Precise PID Temperature Control up to 1700°C
The OTF-1700X-UL is engineered for laboratory applications requiring precise heat treatment of samples within a controlled atmosphere or vacuum environment. It is designed to accommodate a high-quality alumina tube with an outer diameter of 60 mm and length of 1000 mm, enabling processes such as material synthesis, annealing, and sintering at elevated temperatures up to 1700°C.
This split tube furnace features a unique separable design allowing easy access for tube replacement and maintenance. The system includes stainless steel vacuum flanges equipped with needle valves and vacuum gauges, facilitating reliable sealing and monitoring of vacuum or gas flow conditions. The furnace’s temperature is controlled by a PID controller with auto-tune functionality, capable of programming up to 30 heating and cooling segments with ±1°C accuracy.
주요 기능
- Accommodates 60 mm O.D. x 54 mm I.D. x 1000 mm alumina tube; options for 80 mm and 100 mm O.D. tubes available.
- Split-able furnace structure for convenient tube replacement and maintenance.
- Stainless steel vacuum flanges with integrated needle valves and vacuum gauges for controlled atmosphere processing.
- Maximum temperature of 1700°C achievable within 10 minutes under inert gas atmosphere; continuous operation up to 1500°C.
- 16 MoSi2 heating elements (grade 1800) providing uniform heating over a 400 mm zone with a 140 mm constant temperature region.
- PID temperature controller with auto-tune, 30 programmable segments, and ±1°C control accuracy.
- Double-layer steel casing with air cooling to maintain surface temperature below 60°C and high-purity fibrous alumina insulation for energy efficiency.
기술 사양
| 매개변수 | 사양 |
|---|---|
| Alumina Tube Dimensions (O.D x I.D x Length) | 60 mm x 54 mm x 1000 mm (options for 80 mm and 100 mm O.D. available) |
| 최고 기온 | 1700°C (under inert gas, <10 minutes to reach) |
| Continuous Operating Temperature | 1500°C (up to 1600°C for less than 1 hour) |
| Heating Rate | Up to 10°C/min |
| Heating Zone Length | 400 mm total; 140 mm constant temperature zone (±1°C) |
| Heating Elements | 16 pieces MoSi2, grade 1800 |
| Temperature Controller | PID with auto-tune, 30 programmable segments, ±1°C accuracy, DB9 PC communication port (optional PC control module available) |
| Vacuum Sealing | Stainless steel flanges with vacuum gauge, needle valves, 1/4″ pipe and KF25 adapters |
| Vacuum Level Capability | 10^-2 torr (mechanical pump), 10^-5 torr (turbopump), leak rate < 5 mtorr/min |
| Power Supply | 208-240 VAC, single phase, 50/60 Hz, 7 kW (40A breaker) |
대표적인 적용 분야
Ideal for heat treatment processes requiring controlled atmospheres or vacuum, such as sintering ceramics, annealing metals, material synthesis, and research involving high-temperature reactions under inert or vacuum conditions.
구성 및 작동 안내
The furnace includes a pair of fibrous ceramic tube blocks to reduce heat radiation inside the tube; these must be fully inserted before heating. The vacuum pump is not included and should be ordered separately. Optional accessories include PC control modules, digital vacuum gauges, various tube sizes, and vacuum flange fittings. Users must ensure proper gas flow regulation with a two-stage pressure regulator limiting pressure below 3 PSI and avoid opening the furnace lid above 100°C to prevent thermal shock. The alumina tube is designed for vacuum and low-pressure use below 0.02 MPa, and graphite crucibles should not be used inside the alumina tube.