Kategorie

Kontaktieren Sie uns

GSL-1700X-60-VTQ High-Temperature Quenching Tube Furnace with Ø60mm Alumina Tube and Electromagnetic Sample Release

Designed for high-temperature materials research and phase studies.
Features a Ø60ר54 mm alumina tube for controlled atmosphere processing.
Operates up to 1700°C with a 135 mm heating length.
Includes automatic electromagnetic hanger for rapid quenching in liquid media.

Konformität: CE-zertifiziert
Maßgeschneidert: OEM/ODM
Verpackung: Sperrholzkiste

Produkte

GSL-1700X-60-VTQ High-Temperature Quenching Tube Furnace with Ø60mm Alumina Tube and Electromagnetic Sample Release

The GSL-1700X-60-VTQ is engineered for advanced materials research requiring precise high-temperature treatment and rapid quenching capabilities. It is particularly suited for studying phase transformations and microstructural evolution under controlled atmospheres or vacuum conditions. The furnace accommodates samples up to 50 mm in diameter and 65 mm in length within a high-purity alumina tube, ensuring chemical inertness and thermal stability during processing.

This tube furnace integrates a Ø60 mm outer diameter by Ø54 mm inner diameter alumina processing tube with a 135 mm effective heating length. The heating elements consist of four U-shaped molybdenum disilicide (MoSi2) elements rated for operation up to 1700°C, with a constant temperature zone of 60 mm controlled within ±5°C. The furnace structure features double-layer steel casing and high-purity fibrous alumina insulation rated for 1800°C, optimizing thermal efficiency and safety.

Wichtigste Merkmale

  • Maximum operating temperature of 1700°C for short durations; continuous operation between 800°C and 1600°C.
  • Ø60ר54 mm high-purity alumina tube with 850 mm length and included ceramic tube blocks to minimize heat loss.
  • Automatic electromagnetic sample release system enabling rapid quenching by dropping samples into a built-in liquid quenching tank.
  • Vacuum and gas sealing flange assembly with valves and vacuum gauge supporting pressures down to 50 mTorr.
  • Heating rate control with PID temperature controller and optional Eurotherm 3000 series for enhanced precision.
  • Quenching tank integrated below the furnace with manual gate valve and safety interlocks for inert gas or vacuum quenching.
  • Sample size compatibility up to 50 mm diameter and 65 mm length, suitable for crucibles or direct sample hanging.

Technische Daten

Parameter Spezifikation
Maximale Betriebstemperatur 1700°C (for less than 1 hour)
Continuous Operating Temperature Range 800°C to 1600°C
Heating Zone Length 130 mm
Konstanttemperaturzone 60 mm ±5°C
Heizelemente 4 U-shaped MoSi2 elements (1800 grade)
Processing Tube Material and Dimensions 99% High purity alumina (Al2O3), ID 54 mm, OD 60 mm, Length 850 mm
Sample Size Capacity Up to Ø50 mm × 65 mm length
Stromversorgung 208-240 VAC, 30 A, single phase, 50/60 Hz, 4 kVA
Temperaturregler PID with auto-tune, 30 segment programming; optional Eurotherm 3008 for ±0.1°C accuracy
Vakuumfähigkeit Down to 50 mTorr with mechanical pump
Heating/Cooling Rates < 2°C/min at ≥1400°C; < 5°C/min between 1200°C and 1400°C; < 10°C/min below 1200°C
Quenching System Electromagnetic sample release with stainless steel quenching tank and manual gate valve

Typische Anwendungsbereiche

This furnace is ideal for research involving high-temperature phase transformations, sintering studies, and microstructure analysis under controlled atmospheres or vacuum. It supports rapid quenching to preserve transient phases or microstructures formed at elevated temperatures. Applications include ceramics, metallurgy, and advanced materials development.

Hinweise zur Konfiguration und zum Betrieb

The furnace includes a vacuum-tight alumina tube assembly with flanges, valves, and a vacuum gauge. The electromagnetic sample release system automatically drops the sample into the quenching tank upon deactivation. Users must employ ice water or cooled oil as quenching media when operating under vacuum to ensure safety and effective cooling.

A two-stage pressure regulator is required to maintain gas pressure below 3 PSI during inert gas quenching. The furnace should not be operated with graphite crucibles inside the alumina tube. Temperature calibration prior to use is recommended for accurate control. Optional accessories include advanced temperature controllers and PC communication modules for remote monitoring and programming.