Kategorie

Kontaktieren Sie uns

OTF-1200X-HP-III-80-GF Three-Zone High-Pressure Tube Furnace with Gas Flow Control, Nickel-Based Superalloy Vessel, 1100°C Max Temperature

Designed for high-pressure material processing under controlled oxygen or inert gas atmospheres.
Features a nickel-based superalloy pressure vessel with 50 mm inner diameter for durability.
Operates up to 1100°C with three heating zones totaling 450 mm length.
Includes integrated gas flow control system with digital pressure sensors and mass flow controller.

Konformität: CE-zertifiziert
Maßgeschneidert: OEM/ODM
Verpackung: Sperrholzkiste

Produkte

OTF-1200X-HP-III-80-GF Three-Zone High-Pressure Tube Furnace with Gas Flow Control

The OTF-1200X-HP-III-80-GF furnace is engineered for advanced laboratory applications requiring precise high-pressure and high-temperature environments. It is particularly suited for processing iron-based superconducting materials, advanced oxide ceramics, and compact hot isostatic pressing (HIP) research. The system supports controlled atmospheres of oxygen or inert gases with integrated gas flow management to ensure process stability and safety.

This split-type tube furnace incorporates three independently controlled heating zones, each 150 mm in length, providing a total heating zone of 450 mm with a 300 mm constant temperature region. The pressure vessel is constructed from GH747 nickel-based superalloy, offering excellent mechanical strength and thermal stability under high-pressure conditions. The vessel has an 85 mm outer diameter, 50 mm inner diameter, and 1000 mm length, enclosed within a steel safety frame to contain flange assemblies in case of failure.

Wichtigste Merkmale

  • Maximum operating temperature of 1100°C for durations under one hour, with continuous operation at 1000°C.
  • Three precision digital temperature controllers with PID control and programmable segments ensure ±1°C stability.
  • Integrated gas flow control system with three pressure sensors, a high-pressure mass flow controller (max 10 MPa), and adjustable pressure relief valves.
  • Nickel-based GH747 superalloy pressure vessel designed for high-pressure oxygen and inert gas atmospheres.
  • Steel safety frame and copper-tube water-cooling rings with included 16 L/min recirculating water chiller for thermal management.
  • PC control software and included laptop enable remote monitoring and operation.
  • Safety features include over-temperature protection, broken thermocouple detection, and strict gas compatibility restrictions (no hydrogen or flammable gases).

Technische Daten

Parameter Spezifikation
Furnace Type Split-type tube furnace with three heating zones and air-cooling layer
Heating Zone Lengths 150 mm + 150 mm + 150 mm (Total 450 mm)
Konstanttemperaturzone 300 mm ±5°C
Höchsttemperatur 1100°C (less than 1 hour)
Continuous Operating Temperature 1000°C
Maximum Heating Rate ≤10°C/min
Temperaturregelung Three digital PID controllers, 30 programmable segments each, ±1°C stability
Thermoelement OMEGA Type K, 3 mm OD, high-temp connector rated to 220°C
Pressure Vessel Material GH747 nickel-based superalloy
Abmessungen des Schiffes 85 mm OD × 50 mm ID × 1000 mm length
Maximum Operating Pressure 140 bar at 800°C, decreasing to 2.5 bar at 1100°C
Gas Compatibility Oxygen and inert gases only; hydrogen and flammable gases prohibited
Gas Flow Control Integrated mass flow controller (max 10 MPa), digital pressure sensors, relief valves
Kühlsystem Copper-tube water cooling rings with 16 L/min recirculating water chiller included
Stromversorgung AC 208–240 V, 50/60 Hz, max 4 kW

Typische Anwendungsbereiche

This furnace is ideal for research on iron-based superconductors, advanced oxide ceramics, and compact hot isostatic pressing processes. It is also applicable for hydroretorting of oil shale and other high-pressure, high-temperature material synthesis or treatment requiring controlled gas atmospheres.

Hinweise zur Konfiguration und zum Betrieb

The system includes a steel safety frame to contain the pressure vessel and flange assembly. Users must never place samples directly inside the vessel without a crucible or isolation foil to prevent chemical reactions with the vessel alloy. The furnace must not be operated above 1100°C, and pressure limits must be strictly observed as temperature increases.

Hydrogen and flammable gases are strictly prohibited. A two-stage pressure regulator must be installed on gas cylinders to limit supply pressure below 3 PSI. Optional accessories include a torque wrench for flange bolts, alumina testing fixtures, and high-pressure electrical feedthroughs for advanced electrical property measurements under high-pressure conditions.

Produktbezogene Informationen

MSK-SP-01A laboratory product view 1
PC22-UV laboratory product view 1
MSK-SFM-11-LD_01
1
Zum Seitenanfang scrollen