GSL-KR200 Flammable Gas Control and Disposal Device with Hydrogen Leak Detection and Flow Control
The GSL-KR200 is engineered to manage and safely dispose of residual flammable gases generated during laboratory processes, particularly in setups involving reducing atmospheres such as hydrogen flow. It is designed to integrate seamlessly with MTI tube and atmosphere furnaces, providing a reliable solution for controlling exhaust gas hazards.
This device employs a self-ignition heating coil to combust residual flammable gases, preventing their accumulation and potential hazards. It also incorporates a hydrogen gas detector that monitors for leaks and triggers an automatic shutdown of gas flow and furnace heating if hydrogen concentrations exceed 20% of the lower explosive limit (LEL). The system includes safety interlocks and flow control components to ensure secure operation.
Wichtigste Merkmale
- Self-ignition heating coil for combustion of residual flammable gases along exhaust lines.
- Compatible with various MTI furnaces and designed for reducing environment experiments.
- Integrated hydrogen gas detector with preset alarm at 20% LEL to ensure safety.
- Automatic interlocks that cut off gas flow and furnace heating upon emergency detection.
- Dual solenoid valves allowing selection between inert and flammable gas flow paths.
- One float flow meter providing flow rate control from 100 to 1000 sccm (air equivalent).
- Check valve to prevent backflow of gases, enhancing operational safety.
Technische Daten
| Parameter | Spezifikation |
|---|---|
| Leistung | 200 W |
| Spannung | 110 V, 208-240 VAC, single phase, 50-60 Hz |
| Flow Rate | 100-1000 sccm (air, via float flow meter) |
| Gas Inlets | Two 1/4″ compression tube fittings (for inert and flammable gases) |
| Hydrogen Gas Detector | External connection; alarm preset at 20% LEL; triggers automatic shutdown |
| Safety Features | Interlocks for gas flow and furnace heating cutoff; check valve to prevent backflow |
Typische Anwendungsbereiche
The GSL-KR200 is ideal for laboratory experiments requiring controlled reducing atmospheres, such as hydrogen flow studies, where safe disposal of residual flammable gases is critical. It is suitable for integration with MTI tube and atmosphere furnaces used in materials research and chemical processing.
Configuration and Operating Notes
The device requires installation of a metallic mesh cover at both ends of the furnace tube when used with tube furnaces to enhance safety in emergencies. It must be operated in well-ventilated areas or fume hoods. The hydrogen gas detector must be properly installed and calibrated. Operators should receive training from qualified personnel and adhere strictly to laboratory safety protocols. Regular leak inspections and system purging before use are strongly recommended to maintain safe operation.
Maße & Versand
| Product dimensions | 360L × 340W × 400H (mm) |
|---|